INSTALLATION
Milli-Q
A
c
G
r
B
io
S
yn
A10
12
Installation of the Q-Gard purification pack (Figure 4)
(The Q-Gard is only used on Milli-Q systems equipped with a pack adapter (A))
1. Raise the pack adapter (A) to its highest position. Remove the two protective plastic inserts (B)
on the pack adapter.
2. Remove the two protective inserts on the Q-Gard pack (C). Wet the two O-rings on the Q-Gard
with pure water.
3. Push the Q-Gard so that the pack adapter metal rod (D) goes through the hole at the top of the
Q-Gard.
- Lift the Q-Gard slightly and push the bottom of it into the small opening (E) on the system.
- Push the top of the Q-Gard until it is secure.
4. Lock the Q-Gard in place with the metal locking clip (F) on the end of the metal guide pin (D).
5. Bring the adapter cap down to its lowest position (G) so that it covers the top of the Q-Gard.
Electrical connection of system
1. Connect the power cord (figure 2, B) to the Milli-Q (figure 3, L). Connect the other end of the
power cord to an earth grounded power source.
2. Check that the point of use Milli-Q gun trigger is in the upright position.
3. Turn on the system electrical switch (figure 1, H) by putting it to position I.
4. The serial number is displayed for 10 seconds. Please record the type of system and serial
number below.
Example:
SR. N°
Note: Contact Millipore Technical Service when connecting a low level tank sensor to the Milli-Q
system. As an option, it is possible to get a “NO FEED WATER” message when there is no water in
the reservoir.
Installation of the Quantum Ultrapure Cartridge (Figure 4)
Note: It is important to have the system electrical power switched on before installing the Quantum
Ultrapure Cartridge.
1. Open the blue door on the front of the Milli-Q by pressing the 2 latches (H) on the right side of
the door to open it.
2. Remove the two protective inserts from the Quantum cartridge. Wet the two rubber O-rings on
the Quantum cartridge with pure water.
3. Install the Quantum cartridge and push it in as far as it will go.
4. Close the door. It is necessary to fully snap the latches (H) shut to hold the Quantum cartridge
inside.
GRADIENT VI.03
SR.N° F8 KM35924