Troubleshooting messages (continued)
STATUS/PROBLEM CAUSE REMEDY
EXCHANGE
UV LAMP
START SANIT.
A10 ERROR #
EXCHANGE
A10 UV
A10 CLEANING
The limit of the UV lamp life has
been reached.
A cleaning cycle of the UF
cartridge needs to be started.
An error has occurred with the
A10 TOC module.
The limit of the UV lamp life has
been reached.
The Milli-Q has the A10 inside.
The cartridges were just replaced
or an A10 cleaning was started
from the SERVICE MENU.
Replace the UV lamp. Call Millipore
Technical Service.
Start a cleaning procedure.
See the MAINTENANCE chapter.
Press OPERATE/STANDBY to reinitialise
the Milli-Q. If the error persists, then
contact Millipore Technical Service .
Replace the UV lamp in the A10.
Contact Millipore technical Service .
Let the Milli-Q finish the 60 minute A10
cleaning cycle. The Milli-Q can dispense
water during this mode.