Millipore Milli-Q Plus Specifications Page 28

  • Download
  • Add to my manuals
  • Print
  • Page
    / 173
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 27
APPENDIX 1
Milli-Q
A
c
G
r
B
io
S
yn
A10
29
Interrupting the sanitisation cycle of a UF cartridge
If a sanitant chemical has been introduced into the Milli-Q, it is absolutely necessary to let the sanitisation
cycle finish in order to completely rinse the sanitant from the system.
If no sanitant has been introduced, then the sanitisation cycle can be cancelled by following these instruc-
tions.
Press CLEANING for 10 seconds
At the end of the sanitisation cycle, the
Milli-Q will go back to its initial mode of
operation.
Purging trapped air from the UF cartridge
In either PRE OPERATE or PRODUCT
Press MENU for 2 seconds
Press MENU 4 times
Open the point of use valve to begin the
AIR PURGE cycle.
Dispense product water from the system to
the drain.
At the end of the air purge, the system will
go back into STANDBY mode.
Close the point of use valve to place the
system in PRE OPERATE mode.
Interrupting the A10 cleaning cycle
If an A10 cleaning cycle has been started from the MENU service mode, then it can be cancelled at any
time by pressing on the MENU button.
OPERATE MEASURE CLEANING MENU
STAND/BY
OPERATE MEASURE CLEANING MENU
STAND/BY
OPERATE MEASURE CLEANING MENU
STAND/BY
PRE OP ERATE
AIR PURGE 5mn
SERVIC E: UF
AIR P URGE
PROD. TIM E SETUP
COUNTER: 0mn
STA NDB Y
SAN.CYCLE 420mn
STA NDB Y
CLOSE THE VALVE
Page view 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 172 173

Comments to this Manuals

No comments