Installing the Q-Gard Pack
Procedure
Follow the steps below to install a new Q-Gard Pack.
Step Action Diagram
1 Start in STANDBY Mode.
NOTE:
The Q-GARD PACK OUT
Alarm message is not shown at
this time. By following the
instructions earlier in this
manual, the alarm was
cancelled.
2 Open the left door of the
System Cabinet.
Remove the 2 protective caps
located on the ports inside.
3 Remove the covers on the 2
ports of the Q-Gard Pack.
Make sure the rubber O-rings
are firmly in place.
Wet the O-rings with water.
4 Push the top of the Q-Gard
Pack into the ports on the
System.
Continued on next page