- 65 -
Replacing the Quantum Cartridge, Continued
Rinsing
The Quantum Cartridge, when newly installed, needs to be rinsed. This
ensures optimal water quality.
Step Action Diagram
1 Locate the clear tubing and the
barbed fitting from the System
accessories bag.
Screw the barbed fitting onto
the POD Unit.
NOTE:
Do not use any white tape on
the threads of the barbed fitting.
An O-ring is located inside the
POD Unit.
Push one end of the clear
tubing onto the end of the
barbed fitting.
Place the other end of the clear
tubing into a sink.
2 The System must be in READY
Mode.
3 Push the plunger down on the
POD Unit.
In a few minutes, water should
dispense from the
POD Unit.
Continued on next page
Comments to this Manuals